LMK mobile R
High quality system camera for luminance photography
The application of high-quality digital cameras in the LMK-system allows an easy and fast capture of luminance images. For evaluating the images, the complete functionality of the LMK LabSoft software can also be utilized.



Technical data
Sensor
CMOS Canon 35 mm format
6249 (H) x 4160 (V) Resolution of the image receiver
3132 (H) x 2090 (V) Resolution of the luminance image
14 bit CR3 image files with RGB layer structure, uncompressed
Dynamic range:
Single picture measurement (SinglePic)
[ 1:4000 ]
High Dynamic measurement (AEB)
[ 1:30000 (1/1000 sec. < ti < 8 sec.) ]
Data interface:
USB 3.0/Wi-Fi and/or SD/SDHC memory card
Focal Length / Visual field angle:
- Focal length 35 mm: 57°(H) x 38°(V)
Selection of measuring range
Selecting aperture, exposure time and ISO speed
Metrological specification
Spectral matching
Numerical transformation of R,G,B - sensor data
Calibration uncertainty
∆L [ 2.5% ]
Uniformity
∆L [ 5.1% ]
Focus uncertainty
∆L [ 2.5% ]
Repeatability
∆L [ < 1.3% ]
Restrictions
- Can not be used for measuring coloured light sources (i.e. LED)
- Limited use for measuring modulated light source with a high depth of modulation
Result data
- Luminance distributions in measuring images L(x,y)
- Statistical data for being used in calculation programs (z.B. EXCEL®, MatLAB®, LabVIEW®)